Data analysis in the chemical laboratory: II. The end-point estimation in instrumental titrations by nonlinear regression
KUPKA, K. MELOUN, M.
Czech title
Stanovení bodu ekvivalence instrumentálních titrací pomocí nelineární regrese
English title
Data analysis in the chemical laboratory: II. The end-point estimation in instrumental titrations by nonlinear regression
Type
journal article - other
Language
en
Original abstract
The regression model for a two-segments titration curve with a break-point at the end-point is analyzed. Both linear and nonlinear shapes of the titration curve segments are treated. An effective and simple method discriminates which of two segments is linear or curved. The point and interval estimates of the end-point are calculated by the nonlinear least squares of curve fitting technique.
Czech abstract
Využití po částech spojitého regresního modelu pro stanovení bodu zlomu pomocí metod nelineární regrese.
English abstract
The regression model for a two-segments titration curve with a break-point at the end-point is analyzed. Both linear and nonlinear shapes of the titration curve segments are treated. An effective and simple method discriminates which of two segments is linear or curved. The point and interval estimates of the end-point are calculated by the nonlinear least squares of curve fitting technique.
Keywords in Czech
Bod změny, po částech spojité modely, titrační křivka
Keywords in English
Change point, piecewise regression, titration curve
Released
01.02.2001
Publisher
Elsevier Science B.V.
ISSN
0003-2670
Volume
429
Number
1
Pages from–to
171–183
Pages count
13
BIBTEX
@article{BUT51007,
author="Karel {Kupka} and Milan {Meloun},
title="Data analysis in the chemical laboratory: II. The end-point estimation in instrumental titrations by nonlinear regression",
year="2001",
volume="429",
number="1",
month="February",
pages="171--183",
publisher="Elsevier Science B.V.",
issn="0003-2670"
}