Original abstract

During the analysis of the fracture surfaces of the destructed hip joint endoprosthesis ceramics heads it was detected that the shape (size and the position) of the cone end in the bottom of the cone opening were very different. The paper presents the influence of five typical shapes of the cone end on the stress in the ceramic head. The computational modelling is realized for two outer diameters of the ceramic head – 28 mm and 32 mm and three depths of it's cone. The system FEM – ANSYS was used for computational modelling of the stress and deformation of analyzed system (testing stem and ceramic head). Loading of the analyzed system is according to norm ISO 7206-5. There are assumed shape deviations of the cone surfaces – interaction of the ovality and different degree of taper of the stem and the head cone.

Czech abstract

Príspevek se zabývá problematikou analýzy napetí u ruzných tvarových typu prechodu kužele do dna u keramických hlavic totální kycelní endoprotézy.

English abstract

During the analysis of the fracture surfaces of the destructed hip joint endoprosthesis ceramics heads it was detected that the shape (size and the position) of the cone end in the bottom of the cone opening were very different. The paper presents the influence of five typical shapes of the cone end on the stress in the ceramic head. The computational modelling is realized for two outer diameters of the ceramic head – 28 mm and 32 mm and three depths of it's cone. The system FEM – ANSYS was used for computational modelling of the stress and deformation of analyzed system (testing stem and ceramic head). Loading of the analyzed system is according to norm ISO 7206-5. There are assumed shape deviations of the cone surfaces – interaction of the ovality and different degree of taper of the stem and the head cone.

Keywords in Czech

Kyčelní endoprotéza, napjatostní analýza, keramika, tvar přechodu

Keywords in English

Total hip joint endoprosthesis, ceramics head, shape of the cone opening, stress, computational modelling.

RIV year

2008

Released

06.12.2008

Publisher

Springer

Location

Singapore

ISBN

978-3-540-92840-9

Book

13th International Conference on Biomedical Engineering

Pages from–to

2012–2015

Pages count

4

BIBTEX


@inproceedings{BUT29087,
  author="Vladimír {Fuis} and Jozef {Varga},
  title="Stress Analysis of the Hip Joint Endoprosthesis Ceramics Head with Different Shapes of the Cone Opening",
  booktitle="13th International Conference on Biomedical Engineering",
  year="2008",
  month="December",
  pages="2012--2015",
  publisher="Springer",
  address="Singapore",
  isbn="978-3-540-92840-9"
}