Publication detail
Reliability of the Ceramic Head of the Total Hip Joint Endoprosthesis Using Weibull's Weakest-link Theory
FUIS, V. NÁVRAT, T. HLAVOŇ, P. JANÍČEK, P.
Czech title
Spolehlivost keramické hlavice totální kyčelní endoporotézy určená pomocí Weibullovy teorie nejslabšího článku
English title
Reliability of the Ceramic Head of the Total Hip Joint Endoprosthesis Using Weibull's Weakest-link Theory
Type
conference paper
Language
en
Original abstract
The paper deals with calculation of the parameters of ceramic material from a set of destruction tests of ceramic heads of total hip joint endoprosthesis. The standard way of calculation of the material parameters consists in carrying out a set of 3 or 4 point bending tests of specimens cut out from parts of the ceramic material to be analysed. In case of ceramic heads, it is not possible to cut out specimens of required dimensions because the heads are too small (if the cut out specimens were smaller than the normalised ones, the material parameters derived from them would exhibit higher strength values than those which the given ceramic material really has). On that score, a special testing jig was made, in which 40 heads were destructed. From the measured values of circumferential strains of the head's external spherical surface under destruction, the state of stress in the head under destruction was established using the final elements method (FEM). From the values obtained, the sought for parameters of the ceramic material were calculated using Weibull's weakest-link theory.
Czech abstract
Příspěvek se zabývá určením materiálových parametrů keramiky pomocí destrukčních zkoušek keramických hlavic totální kyčelní endoprotézy a jejího následného stochastického zpracování.
English abstract
The paper deals with calculation of the parameters of ceramic material from a set of destruction tests of ceramic heads of total hip joint endoprosthesis. The standard way of calculation of the material parameters consists in carrying out a set of 3 or 4 point bending tests of specimens cut out from parts of the ceramic material to be analysed. In case of ceramic heads, it is not possible to cut out specimens of required dimensions because the heads are too small (if the cut out specimens were smaller than the normalised ones, the material parameters derived from them would exhibit higher strength values than those which the given ceramic material really has). On that score, a special testing jig was made, in which 40 heads were destructed. From the measured values of circumferential strains of the head's external spherical surface under destruction, the state of stress in the head under destruction was established using the final elements method (FEM). From the values obtained, the sought for parameters of the ceramic material were calculated using Weibull's weakest-link theory.
Keywords in English
FEM, Weibull weakest link theory, destruction, ceramics, material parameters
RIV year
2006
Released
27.08.2006
Publisher
Springer
Location
Seoul
ISBN
3-540-36839-6
Book
World Congress on Medical Physics and Biomedical engineering 2006
Pages count
4
BIBTEX
@inproceedings{BUT24128,
author="Vladimír {Fuis} and Tomáš {Návrat} and Pavel {Hlavoň} and Přemysl {Janíček},
title="Reliability of the Ceramic Head of the Total Hip Joint Endoprosthesis Using Weibull's Weakest-link Theory",
booktitle="World Congress on Medical Physics and Biomedical engineering 2006",
year="2006",
month="August",
publisher="Springer",
address="Seoul",
isbn="3-540-36839-6"
}