Publication detail
Cyclic Fatigue of Dental NiTi Instruments after Plasma Nitriding
BUMBÁLEK, M. JOSKA, Z. POKORNÝ, Z. SEDLÁK, J. MAJERÍK, J. NEUMANN, V. KLÍMA, K.
Czech title
Cyklická únava zubních NiTi nástrojů po plazmové nitridaci
English title
Cyclic Fatigue of Dental NiTi Instruments after Plasma Nitriding
Type
journal article in Web of Science
Language
en
Original abstract
This study investigated the possibility of nitride NiTi instruments using low-temperature plasma nitriding technology in a standard industrial device. Changes in the properties and fatigue life of used NiTi instruments before and after low-temperature nitriding application were investigated and compared. Nontreated and two series of plasma-nitrided NiTi instruments, designed by Mtwo company with tip sizes of 10/.04 taper, 15/.05 taper, and 20/.06 taper, were experimentally tested in this study. All these instruments were used and discarded from clinical use. The instruments were tested in an artificial canal made of stainless steel with an inner diameter of 1.5 mm, a 60° angle of curvature, and a radius of curvature of 3 mm. A low-temperature plasma nitriding process was used for the surface treatment of dental files using two different processes: 550 °C for 20 h, and 470 °C for 4 h. The results proved that it is possible to nitride dental instruments made of NiTi with a low-temperature plasma nitriding process. Promising results were achieved in trial testing by NiTi instruments nitrided at a higher temperature. Plasma-nitrided files were found to have, in some cases, significantly higher values than nontreated files in terms of fatigue life. The results showed that the nitriding process offers promising possibilities for suitably modified surface properties and quality of surface layer of NiTi instruments. Within the limitations of the present study, the cyclic fatigue life of plasma-nitrided NiTi dental files can be increased using this surface technology.
Czech abstract
Tato studie zkoumala možnost nitridových NiTi nástrojů využívajících nízkoteplotní plazmovou nitridační technologii ve standardním průmyslovém zařízení. Byly zkoumány a porovnávány změny vlastností a únavové životnosti použitých NiTi nástrojů před a po aplikaci nitridace při nízkých teplotách.
English abstract
This study investigated the possibility of nitride NiTi instruments using low-temperature plasma nitriding technology in a standard industrial device. Changes in the properties and fatigue life of used NiTi instruments before and after low-temperature nitriding application were investigated and compared. Nontreated and two series of plasma-nitrided NiTi instruments, designed by Mtwo company with tip sizes of 10/.04 taper, 15/.05 taper, and 20/.06 taper, were experimentally tested in this study. All these instruments were used and discarded from clinical use. The instruments were tested in an artificial canal made of stainless steel with an inner diameter of 1.5 mm, a 60° angle of curvature, and a radius of curvature of 3 mm. A low-temperature plasma nitriding process was used for the surface treatment of dental files using two different processes: 550 °C for 20 h, and 470 °C for 4 h. The results proved that it is possible to nitride dental instruments made of NiTi with a low-temperature plasma nitriding process. Promising results were achieved in trial testing by NiTi instruments nitrided at a higher temperature. Plasma-nitrided files were found to have, in some cases, significantly higher values than nontreated files in terms of fatigue life. The results showed that the nitriding process offers promising possibilities for suitably modified surface properties and quality of surface layer of NiTi instruments. Within the limitations of the present study, the cyclic fatigue life of plasma-nitrided NiTi dental files can be increased using this surface technology.
Keywords in Czech
nikl titan, endodontický, plazmová nitridace, cyklická únava, zlomenina
Keywords in English
nickel titanium, endodontic, plasma nitriding, cyclic fatigue, fracture
Released
23.04.2021
Publisher
MDPI
Location
Basel
ISSN
1996-1944
Volume
14
Number
9
Pages from–to
1–11
Pages count
11
BIBTEX
@article{BUT171596,
author="Michal {Bumbálek} and Zdeněk {Joska} and Zdeněk {Pokorný} and Josef {Sedlák} and Jozef {Majerík} and Vlastimil {Neumann} and Karel {Klíma},
title="Cyclic Fatigue of Dental NiTi Instruments after Plasma Nitriding",
year="2021",
volume="14",
number="9",
month="April",
pages="1--11",
publisher="MDPI",
address="Basel",
issn="1996-1944"
}