Two-dimensional correction of data measured using a large pressure sensor
POHANKA, M.
Czech title
Dvou rozměrová korekce naměřených dat s použitím tlakového čidla
English title
Two-dimensional correction of data measured using a large pressure sensor
Type
conference paper
Language
en
Original abstract
This paper describes how to obtain precise results using a real finite.size pressure sensor. In some cases, the impact distribution is very small or at least very narrow. Measuring such distributions with a pressure sensor, which is not small enough, results in incorrect data. The pressure is being averaged during the measurement and the shape of the obtained spray distribution is larger and the measured pressure maximum is lower than the real one. This paper describes how to correct the measured spray distribution. A two-dimensional fast Fourier Transformation (FFT) is used to convert both the measured data and pressure sensor to the frquency domain. The correction of the measured spray distribution is done in the frequency domain and then an inverse two-dimensional FFT is used to convert the data back to the space domain. This allows suppressing noise in the measured data that would be extremely amplified. Very good results are presented using both the artificial and real data.
Czech abstract
Tento článek popisuje,jak obdržet přesné hodnoty s použitím tlakového sensoru a jak opravit měřené hodnoty rozložení ostřiku. Dvou rozměrná FFT Fast Fourier Transformation – Fourierova transformace je použita k přepočtení naměřených dat a tlakového čidla na kmitočtovou doménu. Korekce měřeného rozložení ostřiku je provedena v kmitočtové doméně a poté je inverzní dvou rozměrná FFT použita pro přepočet dat zpět do prostorové domény
English abstract
This paper describes how to obtain precise results using a real finite.size pressure sensor. In some cases, the impact distribution is very small or at least very narrow. Measuring such distributions with a pressure sensor, which is not small enough, results in incorrect data. The pressure is being averaged during the measurement and the shape of the obtained spray distribution is larger and the measured pressure maximum is lower than the real one. This paper describes how to correct the measured spray distribution. A two-dimensional fast Fourier Transformation (FFT) is used to convert both the measured data and pressure sensor to the frquency domain. The correction of the measured spray distribution is done in the frequency domain and then an inverse two-dimensional FFT is used to convert the data back to the space domain. This allows suppressing noise in the measured data that would be extremely amplified. Very good results are presented using both the artificial and real data.
Keywords in Czech
tlakové čidlo, dvou rozměrová korekce, frekvenční doména
Keywords in English
pressure sensor,two-dimensional correction,frequency domain
RIV year
2004
Released
23.05.2003
Publisher
Witpress
Location
Londýn
ISBN
1-85312-969-0
ISSN
0257-9731
Book
Computational Methods and Experimental Measurements XI
Journal
Journal of the Chinese Society of Mechanical Engineers
Volume
4
Edition number
1
Pages from–to
86–84
Pages count
9
BIBTEX
@inproceedings{BUT13512,
author="Michal {Pohanka},
title="Two-dimensional correction of data measured using a large pressure sensor",
journal="Journal of the Chinese Society of Mechanical Engineers",
booktitle="Computational Methods and Experimental Measurements XI",
year="2003",
volume="4",
month="May",
pages="86--84",
publisher="Witpress",
address="Londýn",
isbn="1-85312-969-0",
issn="0257-9731"
}