Detail publikace
Napjatostní analýza keramické hlavice totální kycelní endorpotézy s mikro a makro tvaovými odchylkami
FUIS, V. HLAVOŇ, P.
Český název
Napjatostní analýza keramické hlavice totální kycelní endorpotézy s mikro a makro tvaovými odchylkami
Anglický název
Stress Analysis of the Ceramics Head with Micro and Macro Shape Deviations of the Contact Areas
Typ
článek ve sborníku ve WoS nebo Scopus
Jazyk
en
Originální abstrakt
The paper deals with the problems of ceramic head of hip joint endoprosthesis destructions in vivo, and with assessing the impact of shape deflections of conical surfaces on the tensile stress in the head. There are assumed two material characteristics of the stem – the first is linear elastic and the second is bilinear plastic modelled austenitic steel used in the bio-implants. Concerned are shape deviations from the ideal conical surfaces of the stem and the head of the endoprosthesis. The shape deviations may be modelled at the macro-level – this concerns model shape inaccuracies such as deviation from the nominal degree of taper, at the micro-level – when the stochastic distribution of unevenness on the contact areas is respected. The problem of stress in ceramic heads was solved using the algorithm of the finite element method – system ANSYS under ISO 7206-5 loading. In the paper are presented and analysed the results of solution of the macro-level shape deviations and micro-level shape deviations, obtained from measurements made on the cones of stems and heads.
Český abstrakt
Príspevek se zabývá vlivem materiálu dríku na napjatost v hlavici totální kycelní endoprotézy s mikro a makro tvarovými odchylkami.
Anglický abstrakt
The paper deals with the problems of ceramic head of hip joint endoprosthesis destructions in vivo, and with assessing the impact of shape deflections of conical surfaces on the tensile stress in the head. There are assumed two material characteristics of the stem – the first is linear elastic and the second is bilinear plastic modelled austenitic steel used in the bio-implants. Concerned are shape deviations from the ideal conical surfaces of the stem and the head of the endoprosthesis. The shape deviations may be modelled at the macro-level – this concerns model shape inaccuracies such as deviation from the nominal degree of taper, at the micro-level – when the stochastic distribution of unevenness on the contact areas is respected. The problem of stress in ceramic heads was solved using the algorithm of the finite element method – system ANSYS under ISO 7206-5 loading. In the paper are presented and analysed the results of solution of the macro-level shape deviations and micro-level shape deviations, obtained from measurements made on the cones of stems and heads.
Klíčová slova česky
Biomechanika, kyčelní endorpotéza, tvarové odchylky
Klíčová slova anglicky
Biomechanics, hip joint endorporsthesis, shape deviations
Rok RIV
2008
Vydáno
29.09.2008
Místo
Praha
ISBN
978-80-01-04163-5
Kniha
Human biomechanics 2008
Strany od–do
1–2
Počet stran
2
BIBTEX
@inproceedings{BUT30124,
author="Vladimír {Fuis} and Pavel {Hlavoň},
title="Stress Analysis of the Ceramics Head with Micro and Macro Shape Deviations of the Contact Areas",
booktitle="Human biomechanics 2008",
year="2008",
month="September",
pages="1--2",
address="Praha",
isbn="978-80-01-04163-5"
}