Detail publikace

A numerical study of hemodynamic effects on the bileaflet mechanical heart valve

ZBAVITEL, J. FIALOVÁ, S.

Anglický název

A numerical study of hemodynamic effects on the bileaflet mechanical heart valve

Typ

článek ve sborníku ve WoS nebo Scopus

Jazyk

en

Originální abstrakt

The work is focused on calculating hemodynamically negative effects of a flow through bileaflet mechanical heart valves (BMHV). Open-source FOAM-extend and cfMesh libraries were used for numerical simulation, the leaflet movement was solved as a fluid-structure interaction. A real model of the Sorin Bicarbon heart valve was employed as the default geometry for the following shape improvement. The unsteady boundary conditions correspond to physiological data of a cardiac cycle. It is shown how the modification of the shape of the original valve geometry positively affected the size of backflow areas. Based on numerical results, a significant reduction of shear stress magnitude is shown. The outcome of a direct numerical simulation (DNS) of transient flow was compared with results of low-Reynolds URANS model k-omega SST. Despite the limits of the two-dimensional solution and Newtonian fluid model, the suitability of models frequently used in literature was reviewed. Use of URANS models can suppress the formation of some relevant vortex structures which may affect the BMHV's dynamics. The results of this analysis can find use in optimizing the design of the mechanical valve that would cause less damage to the blood cells and lower risk of thrombus formation.

Anglický abstrakt

The work is focused on calculating hemodynamically negative effects of a flow through bileaflet mechanical heart valves (BMHV). Open-source FOAM-extend and cfMesh libraries were used for numerical simulation, the leaflet movement was solved as a fluid-structure interaction. A real model of the Sorin Bicarbon heart valve was employed as the default geometry for the following shape improvement. The unsteady boundary conditions correspond to physiological data of a cardiac cycle. It is shown how the modification of the shape of the original valve geometry positively affected the size of backflow areas. Based on numerical results, a significant reduction of shear stress magnitude is shown. The outcome of a direct numerical simulation (DNS) of transient flow was compared with results of low-Reynolds URANS model k-omega SST. Despite the limits of the two-dimensional solution and Newtonian fluid model, the suitability of models frequently used in literature was reviewed. Use of URANS models can suppress the formation of some relevant vortex structures which may affect the BMHV's dynamics. The results of this analysis can find use in optimizing the design of the mechanical valve that would cause less damage to the blood cells and lower risk of thrombus formation.

Klíčová slova anglicky

numerical study, hemodynamic effects, heart valve

Vydáno

28.06.2019

Nakladatel

E D P SCIENCES

Místo

CEDEX A

ISSN

2100-014X

Kniha

EFM18 – Experimental Fluid Mechanics 2018

Ročník

213

Strany od–do

1–5

Počet stran

5

BIBTEX


@inproceedings{BUT170663,
  author="Simona {Fialová},
  title="A numerical study of hemodynamic effects on the bileaflet mechanical heart valve",
  booktitle="EFM18 – Experimental Fluid Mechanics 2018 ",
  year="2019",
  volume="213",
  month="June",
  pages="1--5",
  publisher="E D P SCIENCES",
  address="CEDEX A",
  issn="2100-014X"
}