Vliv geometrie a topografie třecích povrchů umělých náhrad kyčelního kloubu na součinitel tření

CHOUDHURY, D. VRBKA, M. MAMAT, A.B. STAVNESS, I. ROY, CH.K. MOOTANAH, R. KŘUPKA, I.

Český název

Vliv geometrie a topografie třecích povrchů umělých náhrad kyčelního kloubu na součinitel tření

Anglický název

The impact of surface and geometry on coefficient of friction of artificial hip joints

Typ

článek v časopise ve Web of Science, Jimp

Jazyk

en

Originální abstrakt

Coefficient of friction (COF) tests were conducted on 28-mm and 36-mm-diameter hip joint prostheses for four different material combinations, with or without the presence of Ultra High Molecular Weight Polyethylene (UHMWPE) particles using a novel pendulum hip simulator. The effects of three micro dimpled arrays on femoral head against a polyethylene and a metallic cup were also investigated. Clearance played a vital role in the COF of ceramic on polyethylene and ceramic on ceramic artificial hip joints. Micro dimpled metallic femoral heads yielded higher COF against a polyethylene cup; however, with metal on metal prostheses the dimpled arrays significantly reduced the COF. In situ images revealed evidence that the dimple arrays enhanced film formation, which was the main mechanism that contributed to reduced friction.

Český abstrakt

Článek se zabývá analýzou součinitele tření kyčelních náhrad s nominálním průměrem 28 a 36 mm, přičemž je zkoumáno několik materiálových kombinací, včetně přítomnosti částic opotřebení v kontaktu. V článku je studován také vliv cílené modifikace topografie třecích povrchů na součinitel tření a tloušťku mazacího filmu.

Anglický abstrakt

Coefficient of friction (COF) tests were conducted on 28-mm and 36-mm-diameter hip joint prostheses for four different material combinations, with or without the presence of Ultra High Molecular Weight Polyethylene (UHMWPE) particles using a novel pendulum hip simulator. The effects of three micro dimpled arrays on femoral head against a polyethylene and a metallic cup were also investigated. Clearance played a vital role in the COF of ceramic on polyethylene and ceramic on ceramic artificial hip joints. Micro dimpled metallic femoral heads yielded higher COF against a polyethylene cup; however, with metal on metal prostheses the dimpled arrays significantly reduced the COF. In situ images revealed evidence that the dimple arrays enhanced film formation, which was the main mechanism that contributed to reduced friction.

Klíčová slova česky

Endoprotéza kyčelního kloubu; Mikro dent; Součinitel tření; Utváření mazacího filmu; Velikost a vůle

Klíčová slova anglicky

Hip joint prosthesis; micro dimple; Coefficient of friction; Lubrication film formation; Size and clearance

Vydáno

01.08.2017

Nakladatel

Elsevier

Místo

Amsterdam, The Netherlands

ISSN

1751-6161

Ročník

72

Číslo

8

Strany od–do

192–199

Počet stran

8

BIBTEX


@article{BUT137567,
  author="Dipankar {Choudhury} and Martin {Vrbka} and Ivan {Křupka},
  title="The impact of surface and geometry on coefficient of friction of artificial hip joints",
  year="2017",
  volume="72",
  number="8",
  month="August",
  pages="192--199",
  publisher="Elsevier",
  address="Amsterdam, The Netherlands",
  issn="1751-6161"
}