Detail publikace

Numerická studie vlivu tvarových odchylek na kontakt totální kyčeklní implantát

KOUKAL, M. FUIS, V. FLORIAN, Z. JANÍČEK, P.

Český název

Numerická studie vlivu tvarových odchylek na kontakt totální kyčeklní implantát

Anglický název

A NUMERICAL STUDY OF EFFECTS OF THE MANUFACTURE PERTURBATIONS TO CONTACTS OF THE TOTAL HIP REPLACEMENT

Typ

článek v časopise - ostatní, Jost

Jazyk

en

Originální abstrakt

In the endoprosthesis surgery there is a large percentage of implant defects, which cause the failure of the whole prosthesis. One kind of the total hip replacement functionality loss is acetabular cup pull-off from pelvis bone. This article is aimed at manufacture perturbations analysis as one of the possible reasons to this kind of failure. Dimension and of geometry (roundness) manufacturing perturbations were analyzed. It was found, that these perturbations affect considered values of contact pressure and frictional moment. Contact pressure and frictional moment are quantities affecting replacement success and durability.

Český abstrakt

Jedním problémem kyčelní endoprotézy je uvolnění jamky z pánevní kosti. Důvodem jsou tvatrové odchylky kulové plochy jamky, které jsou v článku analyzovány a bylo zjistštěno, že výrazně ovlvňují velikost kontaktního tlaku a třecí moment mezi hlavicí a jamkou.

Anglický abstrakt

In the endoprosthesis surgery there is a large percentage of implant defects, which cause the failure of the whole prosthesis. One kind of the total hip replacement functionality loss is acetabular cup pull-off from pelvis bone. This article is aimed at manufacture perturbations analysis as one of the possible reasons to this kind of failure. Dimension and of geometry (roundness) manufacturing perturbations were analyzed. It was found, that these perturbations affect considered values of contact pressure and frictional moment. Contact pressure and frictional moment are quantities affecting replacement success and durability.

Klíčová slova česky

Totální kyčelní endoprotéza, MKP, drsnost, kontaktní tlak, biomechanika, třecí moment

Klíčová slova anglicky

Total hip replacement, FEM, roundness, contact pressure, frictional moment, biomechanics

Rok RIV

2011

Vydáno

29.04.2011

Místo

Brno

ISSN

1802-1484

Ročník

18

Číslo

1

Strany od–do

33–42

Počet stran

10

BIBTEX


@article{BUT76519,
  author="Milan {Koukal} and Vladimír {Fuis} and Zdeněk {Florian} and Přemysl {Janíček},
  title="A NUMERICAL STUDY OF EFFECTS OF THE MANUFACTURE PERTURBATIONS TO CONTACTS OF THE TOTAL HIP REPLACEMENT",
  year="2011",
  volume="18",
  number="1",
  month="April",
  pages="33--42",
  address="Brno",
  issn="1802-1484"
}